Live Chat Software by Kayako |
Knowledgebase: Calibration > Calibration - SPIRE
How do I convert my SPIRE data from Jy/beam to Jy/pixel?
Posted by Mark Kidger, Last modified by Mark Kidger on 04 July 2018 02:06 PM
|
|
Answer supplied by Ivan Valtchanov: To do the conversion, you should use the pipeline beam area as provided in the SPIRE Data Reduction Guide, table 6.7, i.e. (465.35, 831.27, 1804.31) arcsec^2 at (250, 350, 500) microns, in order to convert your image from Jy/beam to Jy/pixel. Within the Herschel Interactive Processing Environment (HIPE): you can use the task, convertImageUnit(): newImage = convertImageUnit(image=my250_image_in_Jy_per_Beam, newUnit='Jy/pixel',beamArea=465.35) | |
|